Energizer ATX612 specifications Code, Setspan, Changing Instrument Settings

Models: ATX612

1 17
Download 17 pages 28.21 Kb
Page 10
Image 10

GO CAL PRESS

25LEL APPLY

To continue full calibration, press (E) and the instrument will display the first sensor to be calibrated along with the calibration gas setting. The message APPLY CAL GAS will scroll across the bottom of the display.

NOTE: If the gas concentration does not match the setting, press the MODE switch to abort calibration. See Section 6.1, SETSPAN to change the calibration gas settings.

NOTE: APPLY CAL GAS means connect the sample tubing to the appropriate calibration gas cylinder with either a ILPM or preferrably a demand flow regulator. Connect the tubing to the gas sample inlet of the ATX612. Turn on gas supply.

Apply the calibration gas. The instrument will wait for five minutes to sense that calibration gas has been applied before aborting and failing calibration. When the sensor detects a gas concentration greater than 50% of the calibration gas value, the display will indicate the current full span value for that sensor. The message CAL IN PROGRESS will scroll across the bottom of the display.

When calibration of the sensor has been successfully completed, the instrument will automatically step to the next sensor to be calibrated and the preceding step will be repeated.

When all sensors have been calibrated successfully the instrument will emit a short beep and the sensor configuration will be shown on the display.

If calibration results in marginal sensor span values, sensor identifiers will flash on the display. Marginal calibration will occur if the sensor full span value is less than 70% of the applied gas concentration. A marginal sensor calibration may be an early warning sign that the sensor will soon need to be replaced.

NOTE: If the sensor full span value is less than 50% of the calibration gas value, calibration will fail and the instrument will immediately return to the real time READING mode. When failed calibration or low sensitivity is indicated, verify that the calibration cylinder has not emptied or that the cylinder expiration date has not passed.

After displaying the sensor configuration, the instrument will automatically return to the real time READING mode.

6.

CHANGING INSTRUMENT SETTINGS

 

 

 

Instrument settings, including alarm values, calibration gas

 

 

 

concentrations and security code, may only be accessed

 

 

 

and changed during the instrument startup sequence.

 

 

 

The (+) and (-) hidden keys are used to set instrument

 

 

 

alarm and calibration values. Refer to page 6, Figure 1 for

 

 

 

the location of the hidden keys.

 

 

 

To access the instrument menus:

 

 

 

• Turn the ATX612 off and back on again.

 

 

 

• When the display shows the warm up timer, press the

 

 

 

plus (+) and minus (-) keys simultaneously.

 

 

 

If the instrument security code has been set to a value

 

 

CODE

0

 

other than “0”, the CODE screen will be displayed along

 

 

 

 

with the scrolling prompt PRESS (+) OR (-) TO SET (E)

PRESS

 

TO ENTER. Use the (+) and (-) keys to input the correct

 

security code value and press (E). When the correct code

 

 

 

has been entered successfully, the instrument will

 

 

 

immediately enter the settings mode.

 

 

 

The settings mode consists of three functions:

 

 

 

SETSPAN

 

 

 

ALARMS

 

 

 

CODE

 

 

 

 

 

The scrolling prompt, PRESS (+) FOR NEXT (E) TO

 

 

 

SELECT appears on each screen. Press the (+) to step

 

 

 

through the list and (E) to select any of the functions.

 

 

 

Pressing the MODE switch at any one of the functions

 

 

 

will cause the instrument to return to the normal

 

 

 

operating mode.

 

 

6.1

SETSPAN

 

 

 

 

 

 

 

 

The SETSPAN function allows the user to set the

SETSPAN

 

combustible sensor monitoring range to either LEL or CH4

PRESS

 

and to set the calibration gas values for the combustible

and toxic sensors.

14

15

Page 10
Image 10
Energizer ATX612 specifications Code, Setspan, Changing Instrument Settings