Energizer ATX612 specifications Maintenance, Replacing Alkaline Battery Cells

Models: ATX612

1 17
Download 17 pages 28.21 Kb
Page 13
Image 13

8. MAINTENANCE

8.1CLEANING

Wipe the outside of the instrument with a soft, clean cloth.

Never use solvents or cleaning solutions of any type.

8.2CHANGING THE BATTERY PACK

To change the ATX612 battery pack:

• Hold the instrument with battery pack facing up.

• Using the maintenance tool provided with the instrument, turn counter-clockwise and loosen the two hex screws in the battery pack.

• Remove the battery pack. (See Section 8.2.1 for instructions of replacing the Nicad battery module)

Place the battery pack on the instrument.

Turn the hex screws clockwise until the screws are tight

and reach the stops. DO NOT OVER TIGHTEN.

 

8.2.1 REPLACING THE NICAD BATTERY MODULE

 

The Nicad cells within the battery pack may be replaced

Battery Pack

when necessary by installing a new Nicad battery module.

Hex Screws

To replace the Nicad battery module:

Nicad Battery

Module

Battery Pack

Figure 4

20Figure 5

Remove the battery pack from the instrument as described in Section 8.2.

Remove the two screws which hold the battery module in place as shown in Figure 4.

Lift the Nicad battery module from the battery pack.

Place the new Nicad module in the battery pack.

Replace the two screws which hold the Nicad module in place.

8.2.2REPLACING ALKALINE BATTERY CELLS

The alkaline battery pack holds 6 C-cell batteries. To replace the alkaline battery cells:

Remove the battery pack from the instrument as described in Section 8.2.

Remove the alkaline cells from the battery pack.

Insert the new cells making sure to observe proper polarity.

NOTE: Proper polarity of the alkaline battery cells is identified by the molded “+” and “-” symbols in the bottom of the battery pack. In addition the positive (+) battery contact is identified by the red polarizing tab.

8.3OPENING THE INSTRUMENT FOR SERVICE

To open the instrument for service:

Remove the battery pack as described in Section 8.2.

Remove the four screws from the bottom of the instrument chassis.

Gently lift the case top from the chassis.

8.4INSTALLING OR CHANGING SENSORS

To change sensors in the ATX612:

Open the instrument as described in Section 8.3.

Remove the three screws which hold the sensor manifold in place.

Gently lift the sensor manifold away from the chassis as shown in Figure 5.

NOTE: It is not necessary to disconnect the sample tubing from the manifold to replace the sensors.

21

Page 13
Image 13
Energizer ATX612 specifications Maintenance, Replacing Alkaline Battery Cells