FS-8700-123 MetOne Particle Counter Serial Driver Manual

Page 17 of 26

4.13.Map Descriptor Example 9 – An alternate way of storing the current record.

This example is similar to the one in example 7 where the current record was read. Here passive Map Descriptors have been added to store specifically labeled count records in specific places. In addition to the normal storage of the response data, the driver examines the label which precedes each count value and looks for a matching Map Descriptor. Examples 7, 8 and 9 can be combined and the current record stored in all 3 ways.

In this example a number of passive Map Descriptors are related to the active RDBC Map Descriptor in that they belong to the same device (Node Names are the same and both the active and the passives are ‘Device’) as well as all being ‘Current Record’ Map Descriptors. When the driver receives a response to the poll sent by the active Map Descriptor, then it also looks for matching passive Map Descriptors. As it looks at the label information for each count info it searches for a Map Descriptor whose Met1_Label matches the count label. In this example those labels are “0.1” , “0.3” , “0.5” , “1.0” , “3.0” , “5.0”, “10.” , “FLO” and “LOC”. If there was a count whose label was different from one of these then the driver would not find a matching Map Descriptor and thus would not perform this additional storage for that count.

Example: If the current record contained this fragment …0.1 000123 LOC 000009 STA 000101 … then the driver would use passive Map Descriptors to store the value 123 at offset 0 in the Data Array, store the value 9 at offset 8 in DA= DA_CURR_D01_LBL and would not perform additional storage of the value 101 because it cannot find a label = “STA” . However all these values would be stored in the Data Array called DA_CURR_01 as described in example 7.

Map_Descriptors

Map_Descriptor_Name ,Scan_Interval ,Data_Array_Name ,Data_Array_Offset ,Length ,DA_Byte_Name ,Function ,Node_Name ,Met1_Function ,Met1_SubDev_or_Port

Rd_CurretVals ,0s ,DA_CURR_D01 ,0 ,20 ,DA_DUMP_D01 ,Rdbc ,MET1 ,Current Record ,Device

Map_Descriptors

Map_Descriptor_Name ,Scan_Interval ,Data_Array_Name ,Data_Array_Offset ,Length ,Function ,Node_Name ,Met1_Function ,Met1_SubDev_or_Port ,Met1_Label

Rd_CurretVals1

,0s

,DA_CURR_D01_LBL ,0

,1

,passive ,MET1

,Current Record ,Device

,0.1

Rd_CurretVals2

,0s

,DA_CURR_D01_LBL ,1

,1

,passive ,MET1

,Current Record ,Device

,0.3

Rd_CurretVals3

,0s

,DA_CURR_D01_LBL ,2

,1

,passive ,MET1

,Current Record ,Device

,0.5

Rd_CurretVals4

,0s

,DA_CURR_D01_LBL ,3

,1

,passive ,MET1

,Current Record ,Device

,1.0

Rd_CurretVals5

,0s

,DA_CURR_D01_LBL ,4

,1

,passive ,MET1

,Current Record ,Device

,3.0

Rd_CurretVals6

,0s

,DA_CURR_D01_LBL ,5

,1

,passive ,MET1

,Current Record ,Device

,5.0

Rd_CurretVals7

,0s

,DA_CURR_D01_LBL ,6

,1

,passive ,MET1

,Current Record ,Device

,10.

Rd_CurretVals8

,0s

,DA_CURR_D01_LBL ,7

,1

,passive ,MET1

,Current Record ,Device

,FLO

Rd_CurretVals9

,0s

,DA_CURR_D01_LBL ,8

,1

,passive ,MET1

,Current Record ,Device

,LOC

FieldServer Technologies 1991 Tarob Court Milpitas, California 95035 USA Web:www.fieldServer.com Tel: 408 262-2299 Fax: 408-262-2269 Toll_Free: 888-509-1970 email: support@fieldServer.com