114, 115, 116, and 117 Calibration Information

Table 3. DMM Performance Tests (cont)

Step

Function

Range

Applied

 

Display Reading

 

 

 

 

 

114[1]

115[1]

116[1]

117[1]

 

 

 

 

18

L

6.000V

0V

 

-0.002 to 0.002

 

 

DC Volts

 

 

 

 

 

 

19

6.000V

5V

 

4.973 to 5.027

 

 

 

 

 

 

 

 

 

 

 

 

20

 

60.00V

50V

 

49.73 to 50.27

 

 

 

 

 

 

 

 

 

21

 

600.0V

600V

 

596.8 to 603.2

 

 

 

 

 

 

 

 

 

22

 

600.0V

-600V

 

-596.8 to -603.2

 

 

 

 

 

 

 

 

 

23

m

600.0 mV

6 mV, 45 Hz

 

5.6 to 6.4

 

 

AC Millivolts

 

 

 

 

 

 

24

600.0 mV

600mV, 1 kHz

 

587.7 to 612.3

 

 

 

 

 

 

 

 

 

 

 

 

25

600.0 mV

10 mV

 

9.7 to 10.3

 

 

DC Millivolts

 

 

 

 

 

 

26

600.0 mV

600mV

 

596.8 to 603.2

 

 

 

 

 

 

 

 

 

 

 

 

27

G Diode

2.000 V

1.9V

N/A

 

1.881 to 1.919

 

 

 

 

 

 

 

 

 

28

E

1000 nF

Open

N/A

0 to 2

0 to 2

0 to 2

 

Capacitance

 

 

 

 

 

 

29

9999 μF

900 μF

N/A

881 to 919

881 to 919

881 to 919

 

 

 

 

 

 

 

 

 

30

LoZ

10.00 μF

9 μF

N/A

 

7.18 to 10.82

 

 

Capacitance

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set calibrator to standby, reconfigure leads, and program for amps output

 

 

 

 

 

 

 

 

 

31

A

10.00 A

10A

N/A

9.87 to

N/A

9.87 to

 

DC Amps

10.13

10.13

 

 

 

 

 

 

 

 

 

 

 

 

 

32

?

6.000 A

5.0A, 45 Hz

N/A

4.922 to

N/A

4.922 to

 

AC Amps

5.078

5.078

 

 

 

 

 

 

 

 

 

 

 

 

 

33

$

600.0 μA

600 μADC

N/A

N/A

593.8 to

N/A

 

DC μamps

606.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

34

$

600.0 μA

600 μAAC, 45 Hz

N/A

N/A

590.7 to

N/A

 

AC μamps

609.3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

35

y

 

Open input

N/A

N/A

0PEn

N/A

 

Temperature

 

 

 

 

 

 

36

 

0.0 °C

N/A

N/A

-1.0 to 1.0

N/A

 

 

 

 

 

 

 

 

 

 

37

 

 

400 °C

N/A

N/A

395.0 to

N/A

 

 

 

405.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

38

x

 

0.5 V, 45 Hz

0.2 to 0.8,

 

0.2 to 0.8,

0.2 to 0.8,

 

 

 

 

AC

N/A

AC

AC

 

 

 

 

Annunciator

Annunciator

Annunciator

 

 

 

 

 

 

 

 

 

On

 

On

On

 

 

 

 

 

 

 

 

39

 

 

0.5 V, 0 Hz

0.2 to 0.8,

 

0.2 to 0.8,

0.2 to 0.8,

 

 

 

 

DC

N/A

DC

DC

 

 

 

 

Annunciator

Annunciator

Annunciator

 

 

 

 

 

 

 

 

 

On

 

On

On

 

 

 

 

 

 

 

 

40

 

 

500 V[2],

489.7 to

N/A

489.7 to

489.7 to

 

 

 

500 Hz

510.3

510.3

510.3

 

 

 

 

 

 

 

 

 

 

 

 

10

Page 10
Image 10
Fluke 116, 115, 117, 114 specifications 000V