114, 115, 116, and 117 Calibration Information

Accuracy Specifications

Function

 

Range

 

Resolution

Accuracy

Model[1]

 

 

 

 

 

± ([% of Reading] + [Counts])

 

DC millivolts

600.0 mV

 

0.1 mV

0.5 % + 2

114, 115, 116, 117

 

6.000

V

 

0.001 V

 

 

 

 

 

DC Volts

60.00

V

 

0.01 V

0.5 % + 2

114, 115, 116, 117

 

600.0

V

 

0.1 V

 

 

 

 

 

 

 

 

 

 

DC, 45 to 500 Hz

 

500 Hz to 1 kHz

 

Auto-V LoZ[2] True-rms

600.0

V

 

0.1 V

2.0 % + 3

 

 

4.0 % + 3

114, 116, 117

 

 

 

 

 

45 to 500 Hz

 

500 Hz to 1 kHz

 

AC millivolts[2] True-rms

600.0 mV

 

0.1 mV

1.0 % + 3

 

 

2.0 % + 3

114, 115, 116, 117

 

6.000

V

 

0.001 V

 

 

 

 

 

AC Volts[2] True-rms

60.00

V

 

0.01 V

1.0 % + 3

 

 

2.0 % + 3

114, 115, 116, 117

 

600.0

V

 

0.1 V

 

 

 

 

 

Continuity

600 Ω

 

 

1 Ω

Beeper on <20 Ω, off >250 Ω; detects opens

114, 115, 116, 117

 

 

 

 

 

or shorts of 500 μs or longer.

 

 

600.0

Ω

 

0.1 Ω

0.9 % + 2

 

 

6.000 kΩ

 

0.001 kΩ

0.9 % + 1

 

Ohms

60.00 kΩ

 

0.01 kΩ

0.9 % + 1

114, 115, 116, 117

600.0 kΩ

 

0.1 kΩ

0.9 % + 1

 

 

 

 

6.000 MΩ

 

0.001 MΩ

0.9 % + 1

 

 

40.00 MΩ

 

0.01 MΩ

1.5 % + 2

 

Diode test

2.000

V

 

0.001 V

0.9 % + 2

115, 116, 117

Capacitance

1000 nF

 

1 nF

1.9 % + 2

115, 116, 117

 

10.00

μF

 

0.01 μF

1.9 % + 2

 

 

100.0

μF

 

0.1 μF

1.9 % + 2

 

 

9999 μF

 

1 μF

100 μF - 1000 μF: 1.9 % + 2

 

 

 

 

 

 

>1000 μF: 5 % + 20

 

Lo-Z Capacitance (Power-

 

1 nF to 500

μF

20% + 2

115, 116, 117

up option)

 

(10% +2

typical)

 

 

 

 

 

 

 

6.000

A

0.001 A

 

 

 

 

 

AC Amps True-rms[2]

10.00

A[4]

0.01 A

 

 

 

 

 

20 A overload for 30

 

1.5 % + 3

115, 117

(45 Hz to 500 Hz)

 

seconds maximum, 10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

minutes rest minimum.

 

 

 

 

 

 

 

6.000

A

0.001 A

 

 

 

 

 

 

10.00

A[4]

0.01 A

 

 

 

 

 

DC Amps

20 A overload for 30

 

1.0 % + 3

115, 117

 

seconds maximum, 10

 

 

 

 

 

 

 

minutes rest minimum.

 

 

 

 

 

 

Temperature (Type K

-40 °C to 400 °C

0.1 °C

1.0 % + 10[5]

116

thermocouple)

-40 °F to 752 °F

0.2°F

1.0 % + 18[5]

 

AC μAmps True-rms[2]

600.0

μA

0.1 μA

1.5 % + 3 (2.5 % + 3 > 500 Hz)

116

(45 Hz to 1 kHz)

 

 

 

 

 

 

 

 

 

 

 

 

 

DC μAmps

600.0

μA

0.1 μA

1.0 % + 2

116

4

Page 4
Image 4
Fluke 115, 117, 116, 114 specifications Accuracy Specifications, Function Range Resolution Accuracy Model1