114, 115, 116, and 117

Calibration Information

Meter Buttons Used in the Calibration Steps

When performing the Calibration Adjustment Procedure, the Meter buttons behave as follows. This may be of help determining why a calibration step is not accepted and for determining the input value without referring to Table 4.

Press and hold K to show the measured value. The measured value is not calibrated so it may not match the input value. This is normal.

Press and hold M to display the required input value.

Press g to store the calibration value and advance to the next step. This button is also used to exit calibration mode after the calibration adjustment sequence is complete.

Press Q to toggle the backlight on and off.

Calibration Adjustment Procedure

To adjust the Meter’s calibration, use the following steps. If the Meter is turned off before completion of the adjustment procedure, the calibration constants are not changed.

1.While holding down K, turn the rotary switch from OFF to Ω function. The Meter should display “ZCAL”.

2.Press g once to see the calibration counter.

3.Press g again to start the password entry. The Meter displays “?>>>”.

4.Press the 4 button password.

5.Press g to go to the first calibration step. The Meter displays “C/01” if the password is correct. If the password is not correct, the Meter emits a double beep, displays “?>>>” and the password must be entered again. Repeat step 4.

6.Apply the input value listed for each calibration adjustment step. For each step, select the rotary switch position and apply the input to the terminals as indicated in the Table 4.

Note

Some adjustment steps require additional wait time after the calibrator settles, as noted in Table 4.

7.After each input value is applied, press g to accept the value and proceed to the next step (C/02 and so forth).

Notes

After pressing g, wait until the step number advances before changing the calibrator source or turning the Meter’s rotary knob. Some adjustment steps can take up to several seconds to execute before moving to the next step.

If the knob is not in the correct position for a given step, the meter will flash the unit annunciators until the knob is put in a valid position. The keys that show the reading and required input values are not allowed until the knob is correct.

Likewise, if the rotary switch is not in the correct position or the measured value is not within the anticipated range of the input value, the Meter will emit a double beep and will not continue to the next step when g is pressed.

8.After the final step, the display shows “End” to indicate that the calibration adjustment is complete. Press g to return to meter mode.

14

Page 14
Image 14
Fluke 116, 115, 117, 114 specifications Meter Buttons Used in the Calibration Steps, Calibration Adjustment Procedure