114, 115, 116, and 117
Calibration Information
VoltAle rt
117 TRUE RMS MULTIMETER
|
| Lo / H |
HOLD | MIN MAX | RANGE |
OFF | Hz | |
LoZ |
| V |
|
| V |
|
| mV |
Volt |
| A Hz |
Alert | A |
V
A COM
<.5
OK
OK
10 A FUSED
erc010f.emf
Figure 1. Fuse Testing
Replacing the Battery and Fuse
XWWarning
To avoid shock, injury, or damage to the Meter:
•Remove test leads from the Meter before opening the case or battery door.
•Use ONLY a fuse with the amperage, interrupt voltage, and speed ratings specified.
Battery
Fuse
erc011f.emf
Figure 2. Battery and Fuse Replacement
To remove the battery door for battery replacement, refer to Figure 2 while performing the following:
1.Remove the test leads from the Meter.
2.Remove the battery door screw.
3.Use the finger recess to lift the door slightly.
4.Lift the door straight up to separate it from the case.
5.The battery fits inside the battery door, which is then inserted into the case, bottom edge first, until it is fully seated. Do not attempt to install the battery directly into the case.
6.Install and tighten battery door screw.
To open the case for fuse replacement, refer to Figure 2 while performing the following: 1. Remove the test leads from the Meter.
6