114, 115, 116, and 117
Calibration Information
NORMAL | AUX |
|
| |
V | RTD |
| SC | |
| A SENSE A | UX V | ||
|
|
| OPE | |
|
|
|
| O |
|
|
|
| UT |
TRIG
STBY
7 8
4 5
.
2
+/- 0
.
1 1 7 T R U E R M S M U L T I M E T E R
VoltAlert
Insert Double Banana | Center Meter top on Hi terminal. | |
| Note that LED is RED | |
|
|
|
| erc013f.emf | |
| Figure 4. VoltAlert Testing |
8.Hold the Meter so that the top is vertically and horizontal centered to the banana plug’s Hi terminal. Verify that the Meter’s beeper is sounding continuously and the red LED at the top of the display lights.
9.Return the calibrator to standby mode and verify the Meter’s beeper is silent and the red LED is off.
Calibration Adjustment
The Meter features
The following sections present the features and Meter pushbutton functions available during the Calibration Adjustment Procedure. Should the Meter fail any of the performance tests, perform the Calibration Adjustment Procedure.
Use the following steps to view the Meter’s calibration counter.
1.While pressing K, turn the rotary switch from OFF to Ω function. The Meter should display “ZCAL”.
2.Press g once to view the calibration counter. For example, “n001”
3.Turn the rotary switch to OFF.
Calibration Adjustment Password
To start the Calibration Adjustment Procedure, the correct
Changing the Password
Use the following steps to change the Meter’s password:
1.While pressing K, turn the rotary switch from OFF to Ω function. The Meter should display “ZCAL”.