A PRTwas calibrated to ITS-90 and its calibration certificate states values for
coefficientsR(273.16K), a4, b4, a8, and b8. Set the 1521 coefficients to the
certificatevalues as follows.
1521Coefficients CertificateValue
R(.01) R(273.16K)
aa8
bb8
c0.0
d0.0
a4 a4
b4 b4
Example2:
A PRTwas calibrated to ITS-90 and its calibration certificate states values for
coefficientsR(273.16K), a5, and b5. Set the 1521 coefficients to the certificate
valuesas follows.
1521Coefficients CertificateValue
R(.01) R(273.16K)
aa5
bb5
c0.0
d0.0
a4 a5
b4 b5
7.8.10 IEC751The IEC-751probe type is for standard PRTs (RTDs) conforming to the
IEC-751 or DIN-43760 standards.This type of probe is generally low cost and
has a wide operating range (–180°C to 420°C or higher), but has limited accu-
racy(0.15°C to 4°C). There are no user programmable coefficients associated
with this probe type.
7.8.11 CVDThe CVD (Callendar-Van-Dusen)type is used with PRTs that have unique cal-
ibrationcoefficients or do not match the IEC-751 specification. This type of