Description Command Example
Response
Format
Response
Example Range
Setprobe type p[robe]=IT[S-90]/
IE[C751]/C[VD]/
Y[SI400]/T[HR]
pr ITS-90,
IEC751,
CVD,YSI400,
THERM
ReadITS-90 R(.01) or CVD R0 r[0] r0 r0:9999.999 r0:100.035
SetITS-90 R(.01) or CVD R0 r[0]=n r0=100.106 1to 900
ReadITS-90 coefficient a/b/c/d/a4/b4 a {a/b/c/d/a4/b4}:
-9.999999E-99
a:-2.8644101E-5
SetITS-90 coefficient a/b/c/d/a4/b4=n a=-2.8644E-5 1.0to 9.9e-18
ReadCVD coefficient al[pha]/de[lta]/be[ta] al {al/de/be}: 9.99999999 al:0.0038123
SetCVD coefficient al[pha]/de[lta]/be[ta]=n al=0.0038512 E+/-18
ReadTHERM thermistor
coefficient
b0/b1/b2/b3 b3 {b0/b1/b2/b3}:
9.999999e-99
b3:-9.944070e-12
SetTHERM thermistor
coefficient
b0/b1/b2/b3=n b3=-9.944E-12 E+/-18
Readprobe wires wi[res] wi wi: 9 wi:4
Setprob e wires wi[res]=2/3/4 wi=4 2,3,4
Readprobe calibration date pc[al] pcal pc:yyyy-mm-dd pc:1999-03-28
Setprobe calibration date pc[al]=yyyy-mm-dd pc=1999-03-28 1999-01-01 to
2100-12-31
Readprobe due date pd[ue] pdue pd: yyyy-mm-dd pd:1999-03-28
Setprobe due date pd[ue]=yyyy-mm-dd pd=1999-03-28 1999-01-01to
2100-12-31
Readfilter fi[lter] fi fi: 99.99999 fi:0.5
Setfilter fi[lter]=n fi=0 0.0 to 60.0
Readlow range calibration *c1 *c1 c1:99.999 c1:-0.0121
Setlow range calibration *c1=n *c1=-0.012 -1.999to 1.999
Readhigh range calibration *c2 *c2 c2:9999.9 c2:-0.9
Sethigh range calibration *c2=n *c2=-0.9 -1980.0to
1980.0
Readmeter calibration date mc[al] mcal mc: yyyy-mm-dd mc:1999-03-28
Setmeter calibration date mc[al]=yyyy-mm-dd mc=1999-03-28 1999-01-01to
2100-12-31
Readmeter due date md[ue] mdue md:yyyy-mm-dd : 1999-03-28
Setmeter due date md[ue]=yyyy-mm-dd md=1999-03-28 1999-01-01to
2100-12-31
Legend: []OptionalCommand data
{}Returns eitherinfo rmation
1521 Handheld Thermometer Readout
User’s Guide