39/41B Users Manual

H

 

Table 2. Status Line

 

 

L

 

 

Hold active (screen frozen). appears in lower left screen

VPA

 

to indicate possible high voltage input that is not sensed

 

during Hold.

 

 

Phase Reference Selection. The volts input or the current

 

 

 

 

inputAPis the reference for determining phase shift. Select

 

 

when measuring current only with voltage test leads not

 

 

connected.

2s

 

Smooth Selection. Smooth is active with averaging times of

 

5s

 

2, 5, 10, or 20 seconds. A higher number signifies a more

 

stable measurement reading.

 

 

~10s

 

 

20s

 

 

OL-V

 

Volts Overload Condition for the selected range (over 600V

 

 

 

if in AUTO).

OL-A

 

Amps Overload Condition for the selected range (exceeds

 

 

 

2V peak if in AUTO). Since the input from the Current Probe

 

 

is 1 mV/A, maximum current input is 2000A pk.

OL-VA

 

Both volts and amps maximum inputs have been exceeded.

 

 

 

 

Note

When an overload occurs, all measurement and computation data must be presumed invalid.

AUTO The Tester switches range automatically to deal with changing volts, amps, or watts readings.

MAN Tester does not switch ranges automatically.

MEMX Memory location (X) last accessed. (X = 1 to 8)

SEND Send is active (prints test reports, Model 41B.)

PRNT Print is active (prints screen dumps, Model 41B.)

20

Page 30
Image 30
Fluke 41B, 39 user manual Status Line