Power-Up Configuration

E+ O

Power-Up Configuration

The Tester has two types of configurations: standard and power-up. The standard configuration, which is programmed into the Tester and can always be retrieved, includes the following settings:

FCN

V

Volts Function selected.

DISP

WAVE

Waveform Screen Mode selected.

THD

%RMS

Total Harmonic Distortion computed as percentage of

 

the total RMS value.

CLAMP

80I-500SThe 80i-500s Current Probe is selected.

 

OTH R specifies any current probe other than the 80i-500s. The Tester

has been calibrated to provide either compensation for the 80i-500s or a flat response for other probes.

PR F V Voltage Phase Reference selected.

For Model 41B, KBAUD 9.6 is the standard configuration for Print or Send, and PSON is the standard configuration. The Tester always selects autoranging at power-up.

The power-up configuration, which is initially the same as the standard configuration, can be changed by making entries from the configuration screen. The Tester sets up to the power-up configuration each time it is turned on. You can change the power-up configuration using the following procedure:

1.Press Ewhile pressing Oto turn the Tester on. The configuration screen appears. (See Figure 17.)

2.EPress <and >to highlight different items on the same line. Press to select the currently highlighted item and proceed to the next

line.