39/41B

Users Manual

Operational Test

Use any of the applications shown in this manual to verify volts, amps, and watts measurability. For a full operational test, refer to the Service Manual (P/N 601044).

If the Screen Flashes or is Blank at Power-Up

If the screen flashes or is blank at power-up, it may be caused by a low contrast setting. If neither or the following procedures works, refer to “If Your Tester Does Not Work” for instructions on returning the unit for repair.

If the instrument flashes a dark test pattern and then appears blank:

Tap C, then hold >down for several seconds. A normal screen display should appear. Adjust to the desired contrast, then tap Cagain to store the setting.

If the screen is blank (no dark flash):

Check the batteries. If the batteries read below 1.1V, replace them. If the screen is still blank, refer to “If Your Tester Does Not Work” for instructions on returning the unit.

If Your Tester Does Not Work

If your Tester does not work, make the following basic checks:

Examine the case for physical damage. If you detect damage, contact a Fluke Service Center. (Refer to the list of service centers at the end of this manual.)

Are you testing a live circuit? Test on a known-live circuit.

Check the batteries, test leads, and Current Probe. If necessary, replace any of these items.

Review applicable parts of this manual to make sure you are operating the Tester correctly.

When measuring current without the voltage test leads connected, you must select APfor the phase reference (see Table 1).

54

Page 64
Image 64
Fluke 41B, 39 user manual If Your Tester Does Not Work