Troubleshooting
Troubleshooting
NOTE: Check all possible remediesbefore disassembling the system.

Color Change Solenoid Valves

NOTE: Refer to ElectricalSchematics, page 36. If the colorc hangevalves are not turning on or off correctly, it
could be caused by one of the following.
Cause Solution
1. Air regulator pressure
set too high or too low.
Check that airpressure is at least 85 psi (0.6 MPa, 6.0 bar). Donot go above 100
psi (0.7 MPa, 7.0 bar).
2. Air or electrical lines
damaged or connections
are loose.
Visually inspect airand electrical lines for kinks, damage, or loose connections.
Service or replace as needed.
3. Solenoid failure. Check the applicable solenoid’s LED; see Color Change Board, page 34. If lit,
proceed with the followingchecks. If not lit, go to Cause 4.
Remove the connector for theapplicable solenoid and measure voltage across
the pins on the board:
In a non-hazardous location, replace thesolenoid if voltage is 24 Vdc.
In a hazardous location, replace the solenoid ifvoltage is between 9–15 Vdc.
Test the valves as explained under Maintenance Screen 4 in your PD2K
Operation manual. Valves should open and close quickly. If the valves actuate
slowly, it could be caused by:
Air pressure to the valve actuators is too low. See Cause 1.
Solenoid is clogged. Make sure the air supply has a 5 micron filter installed.
Something is restricting the solenoidor tubing. Check for air output fromthe
air line for the correspondingsolenoid when the valve is actuated. Clear the
restriction.

32 333282B