DISINFECTION

Disinfecting the Pocket Doppler Probe

sNOTICE: Do not allow any liquid to enter the probe socket while disinfecting or immersing the probe.

1.Clean the exterior surface of the probe as recommend- ed in the previous section.

2.To disinfect the probe, immerse the probe in a solution of Benzalkonium Bromide, 0.5% Chlorhexidine, 2% Glu- taraldehyde, or 75% ethanol.

3.Wipe the probe with a clean, dry, non-abrasive cloth to remove any remaining moisture.

sNOTICE: Do not allow any liquid or cleaning solution to remain on the surface of the probe after cleaning.