11 APPENDIX B, EMC INFORMATION

GUIDANCE AND MANUFACTURER’S DECLARATION

Electromagnetic Emissions—for all Equipment and Systems

Guidance and manufacturer’s declaration— electromagnetic emission

The Grafco Ultrasound Pocket Doppler is intended for use in the electromagnetic environment specified below. Do not use the system in environments which do not comply with the specifications listed below.

Emission

Compliance

Electromagnetic environment–

test

 

guidance

RF emission

Group 1

The Grafco Ultrasound Pocket Doppler

CISPR 11

 

uses RF energy only for its internal function.

 

 

Therefore, its RF emissions are very low

 

 

and are not likely to cause any interference

 

 

in nearby electronic equipment.

RF emission

Class B

The Grafco Ultrasound Pocket Doppler

CISPR 11

 

is suitable for use in all establishments,

 

 

including domestic establishments and

 

 

those directly connected to the public

 

 

low-voltage power supply network that

 

 

supplies buildings used for domestic

 

 

purposes.

Electromagnetic Immunity—for all Equipment and Systems

Guidance and manufacture’s declaration—electromagnetic immunity

The Grafco Ultrasound Pocket Doppler is intended for use in the electromagnetic environment specified below. Do not use the system in environments which do not comply with the specifications listed below.

Immunity

IEC 60601

Compliance

Electromagnetic

test

test level

level

environment—guidance

Electrostatic

±6 kV

±6 kV

Floors should be wood,

discharge

contact ±8

contact ±8

concrete or ceramic tile. If

(ESD) IEC

kV air

kV air

floor is covered with synthetic

61000-4-2

 

 

material, the relative humidity

 

 

 

should be at least 30% RH.

14423-INS-LAB-RevD11

36

Page 36
Image 36