Electromagnetic Immunity—for all Equipment and Systems that are not Life-Supporting

Guidance and manufacturer’s declaration— electromagnetic immunity

The Grafco Ultrasound Pocket Doppler System is intended for use in the electromagnetic environment specified below. Do not use the system in environments which do not comply with the specifications listed below.

Immunity

IEC 60601

Compliance

Electromagnetic

test

test level

level

environment—guidance

Radiated

3 V/m 80

3 V/m

Portable and mobile RF

RF IEC

MHz to 2.5

 

communications equipment

61000-4-3

GHz

 

should be used no closer to any

 

 

 

part of the Grafco Ultrasound

 

 

 

Pocket Doppler, including

 

 

 

cables, than the recommended

 

 

 

separation distance calculated

 

 

 

from the equation applicable to

 

 

 

the frequency of the transmitter.

 

 

 

Recommended separation

 

 

 

distance

 

 

 

80 MHz to

 

 

 

800 MHz

 

 

 

800 MHz to

 

 

 

2.5 GHz

 

 

 

Where P is the maximum output

 

 

 

power rating of the transmitter

 

 

 

in watts (W) according to the

 

 

 

transmitter manufacturer and d

 

 

 

is the recommended separation

 

 

 

distance in meters (m).

 

 

 

Field strengths from fixed RF

 

 

 

transmitters, as determined

 

 

 

by an electromagnetic site

 

 

 

survey,a should be less than

 

 

 

the compliance level in each

 

 

 

frequency range.b

 

 

 

Interference may occur in the

 

 

 

vicinity of equipment marked

 

 

 

with the following symbol:

 

 

 

 

14423-INS-LAB-RevD11

37

Page 37
Image 37
Graham Field 14423-8, 14423-2, 14423-4 user manual Distance, Rf Iec