16
Press “Underbed Light” button (#14) to turn underbed light on or off.
Press “CPR Heart” button (#7) to Level/Flatten the bed horizontally for CPR position.
“Safe Mode” (#15) LED light will glow green automatically once power cord is plugged in.
Green light will blink intermittently for approximately 10 minutes. After 10 minutes of non-
activity Hi/Lo Lock icons on the nurse/staff control panel will show orange (locked).
Press “Head”, “Knee”, or “Chair” and “Key” at the same time to unlock individual
lock out functions.
Press “Hi/Lo Locked” to lock out “Hi/Lo”, “Trendelenberg”,
and “Reverse Trendelenberg” functions.
Press “All Locked” and “Key” at
the same time to unlock all functions.
The LED will be lit green when Safe Mode automatically engages
when plugging in the bed. Press “Safe Mode” and “Key” to disable
the Safe Mode feature.
Press “Hi/Lo Locked” and “Key” at the same time
to unlock all Hi/Lo lock out functions.
1 ALL LOCKED BUTTON
2 HEAD DECK UP BUTTON
3 HEAD DECK DOWN BUTTON
4 KNEE & FOOT DECK UP BUTTON
5 KNEE & FOOT DECK DOWN BUTTON
6 CHAIR POSITION BUTTON
7 CPR/UNDO CHAIR POSITION BUTTON
8 KEY LOCK/UNLOCK BUTTON
9 HI/LO (T/TR) LOCKED BUTTON
10 HI/LO UP BUTTON
11 HI/LO DOWN BUTTON
12 TRENDELENBERG
13 REVERSE TRENDELENBERG
14 UNDERBED LIGHT BUTTON
15 SAFE MODE RELEASE BUTTON
LOCK OUT SINGLE FUNCTIONS
To individually lock out the “Head”, “Knee”, “Chair”,
and “Hi/Lo” functions, press the appropriate top
icon (#2, 3, 4, 10, or 12) button and the “Key”
button at the same time. An orange LED lock
symbol will appear under the related icon.
To “Unlock” any of the individual functions, press
the top icon (#2, 3, 4, 10, or 12) and the “Key” Button
(#8) simultaneously. LED lights will notshow up.
LOCK OUT HI/LO FUNCTIONS
To lockout the functions for raising and lowering
the entire bed and tilting the bed for Trendelenberg
positions, press the “Hi/Lo Locked” (#9) button. Orange
LED lock symbols will appear under the icons.
To “Unlock” all Hi/Lo functions, press the “Hi/Lo
Locked” button (#9) and the “Key” button (#8)
simultaneously. Orange LED lights will notshow up.
LOCK OUT EVERYTHING
To lock out all functions, press the “All Locked”
button (#1). Orange LED lock symbols will appear
under the “Head”, “Knee”, “Chair”, “Hi/Lo”, and
“Trendelenberg/Reverse Trendelenberg” icons.
To “Unlock” all functions, press the “All Locked”
button (#1) and the “Key” button (#8) simultaneously.
Orange LED lights will notshow up.
SAFE MODE:
Green light will show automatically when bed is
plugged in and will blink intermittently during 10
minute “sleep mode”. After 10 minutes of lapsed
time since the last function was used, the LED light
shows green and Hi/Lo Lock icons will show orange.
To manually disablethe function, hold the “Safe
Mode” button (#15) and “Key” button (#8) at the
same time. Green LED light will go out.
LOCK SYMBOLS LOCK SYMBOL
LOCK SYMBOLS
LOCK SYMBOLSLOCK SYMBOLS LOCK SYMBOL
LOCK SYMBOLS
Zenith aps seRies
BED OPERATIONS - NURSE/STAFF CONTROL PANEL