LF1031 ONLY: INSTALLATION OF 1010S005 HYDRAULIC PUMP

Removal of Existing Pump

existing pump

existing pump

upper boom

boom

bracket

 

detach existing pump from upper boom

lower mast

mast bracket

detach existing pump from lower mast

1.Follow the instructions in this section to remove the existing pump, shown at above left. If installing pump for the first time, proceed to Installation of new pump on next page.

2.Detach the top of the existing pump from the upper boom bracket, as shown at above center: Use two 9/16" wrenches to remove the hardware that secures top of pump to upper boom bracket. This will free the pump to rotate downward.

3.Detach the bottom of the existing pump from the lower mast bracket, as shown at above right. Use two 9/16" wrenches to remove the hardware that secures bottom of pump to lower mast bracket. This will free the pump — take care that it does not fall.