LF1041 only: charging the battery
The battery used on the LF1041 has an audible alarm that sounds when the battery charge has been depleted by 50%. Charge the battery as soon as possible upon hearing this alarm.
There is also a light on the hand control pendant that illuminates when it is time to recharge the battery.
Info: This patient lift uses a lead acid battery. To prolong the life of the battery, charge the battery fully for 24 hours before initial use. To increase the longevity of the battery, keep the battery fully charged when not in use. Subsequent battery charging time is approximately four hours.
WARNING: After the alarm sounds, there could still be sufficient battery power left to lift a patient. Be aware that a lifted patient cannot be lowered when the battery is discharged. Refer to the Emergency Procedures section that follows for discharged battery operation.
Charging the battery with
Info: The patient lift will not operate during
The LF1041 lift is now ready for use.
Charging the battery with
The LF1041 lift is now ready for use.
LF1041 only: control box power switch (emergency stop button)
In order to use the lift, the control box power switch must be turned ON. To turn the control box power switch ON, turn the red switch button clockwise, it will pop out to the ON position.
To turn the control box power switch OFF, hit it firmly with the palm of your hand; the switch will lock in the depressed position, stopping all motor function.
To prolong the life of the battery, turn the control box power switch OFF when the lift is not in use.
power switch
control box power switch