Hitachi HV-D27A, HV-D37A operation manual White Balance Adjustment

Models: HV-D27A HV-D37A

1 82
Download 82 pages 3.98 Kb
Page 57
Image 57
White Balance Adjustment

Error message

Procedure

AUTO BLACKBAR

Turn off the color bar.

AUTO BLACKNG

Close the lens iris.

 

Avoid taking an image of subject having high luminance such as the

 

sun, or decrease illumination on the microscope.

 

Carry out AUTO BLACK again. If this message appears in repeated

 

attempts, it is necessary to inspect the inside of the camera. In this

 

case, notify your local Hitachi Denshi sales agent or Hitachi Denshi

 

service office

White Balance Adjustment

Carry out white balance adjustment when the illumination condition (color temperature) is changed. Adjust the white balance when using the camera for the first time or after replacing the lens.

1.In the MENU mode, set up WHITE BAL: MEM.

2.Turn off the MENU screen to select the DIRECT mode.

3.Provide a proper aperture value of lens manually.

4.Put an white object in the subject image, and zoom it up.

5.Hold the AWB button pressed for about 2 seconds for automatic white balance adjustment. With MESSAGE RTN:ON, AUTO WHITE appears. At the end of successful adjustment AUTO WHITE:OK appears.

6.If white balance adjustment cannot be made, any of the following messages will appear. Take a proper procedure according to the error message, and then try white balance adjustment again.

43

Page 57
Image 57
Hitachi HV-D27A, HV-D37A operation manual White Balance Adjustment