MAINTENANCE

COMPONENT LOCATION

TESTING EFFICIENCY OF RO MEMBRANE

NOTE: The RO membrane should be replaced annually or when the removal of Total Dissolved Solids - TDS becomes less than 80% efficient. Contact your local Hobart Service office for TDS efficiency testing of your Reverse Osmosis system.

NOTE: To determine grains of hardness delivered to equipment, divide the measured RO TDS by 17.1.

1.Ensure bypass valve is closed.

2.Measure inlet TDS before water enters RO filtration system.

3.Measure RO TDS through sampling line.

4.Determine efficiency percentage using the equation:

Efficiency Percentage =

Inlet TDS ! RO TDS

x 100

Inlet TDS

 

 

Example: The inlet water to an RO filtration system is sampled and found to have TDS equal to 198 ppm. A water sample is then taken from the output of the RO system and it measures a TDS of 17 ppm. Find the efficiency percentage using the above equation and the gathered data.

Solution:

 

 

 

 

 

Efficiency

 

198 ppm ! 17 ppm

 

181 ppm

 

=

x 100 =

x 100 = 0.914 x 100 = 91.4%

198 ppm

198 ppm

Percentage

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-10-