Table 2: Pump On/Off conditions

Status

OK LED on

OK LED off

Pump on

Drip rate greater than 15 dpm

Drip rate less than 15 dpm

 

O.K. relay contacts are closed

Total lack of oil (OIL LED on)

 

 

O.K. relay contacts are open

 

 

 

Pump off

Drip rate is equal to or greater than 2 dpm

Total lack of oil (OIL LED on)

 

O.K. relay contacts are closed

O.K. relay contacts are open

 

 

 

6.Connect an external time-delay relay of approximately four minutes to the O.K. output. This is necessary to extend the time needed for an OK indication during the transition from the pump- off state to the pump-on state.

7.Connect the oil filter (supplied) to the oil tank (Figure 2).

Important: The oil filter is essential as it prevents blockage in the needle valve.

8.Connect the Dripmaster oil outlet (Figure 7 on page 18, and Figure 2) to the well shaft inlet by means of flexible ¼-inch copper tubing.

Setting the Drip Rate

1.Set the dip switch settings for the Normal (pump on) mode using the drip rate setting toggle switches. (see Setting Normal (pump on) Mode on page 16).

Note: There are 50 drops per 1 cc, and there are 3785.4 cc per gallon.