4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual
Conductivity
Cell
Adapter *
*Automatic Temperature Compensator, when specified by catalog number, is mounted inside adapter.
Cable
a/n 23352
Apply teflon tape to both ends of support tube. Tighten all pipe thread fittings two turns beyond hand tight.
Figure 1-4 4908 Conductivity Cell
| Note: Apply a thin film of Silicone grease | |
Do not nick | to all surfaces of bushing and to | |
diameter of support tube as shown. | ||
outside of | ||
Apply grease | ||
support tube | ||
| (See Note above) | |
| Modify per sketch* | |
Tighten securely to pipe |
| |
| These parts should be | |
|
Slice through one side as shown
*
Compression
Bushing
a/n 23353
Figure 1-5 31074357 Removal Device
4 | 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual | 7/99 |