4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual

Conductivity

Cell

Adapter *

*Automatic Temperature Compensator, when specified by catalog number, is mounted inside adapter.

Cable

a/n 23352

Apply teflon tape to both ends of support tube. Tighten all pipe thread fittings two turns beyond hand tight.

Figure 1-4 4908 Conductivity Cell

 

Note: Apply a thin film of Silicone grease

Do not nick

to all surfaces of bushing and to

diameter of support tube as shown.

outside of

Apply grease

support tube

 

(See Note above)

 

Modify per sketch*

Tighten securely to pipe

 

 

These parts should be

 

hand-tightened only

Slice through one side as shown

*

Compression

Bushing

a/n 23353

Figure 1-5 31074357 Removal Device

4

4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual

7/99

Page 12
Image 12