Specifications

2.Specifications

2.1Specifications for 04973 Series

Parameter

Description

 

 

Cell Constant

04973 Series: 0.01, 0.1, 1.0 and 10 as specified

 

 

Electrode Material

0.01 and 0.1 constant, titanium

 

1.0 and 10 constant, high density graphite with

 

Teflon sheath

Maximum Pressure Limit

1724 kPa (250 psig) at rated temperature

 

 

Maximum Continuous Temperature Limit

140°C (284°F) at rated pressure

 

For PVC wire:

 

105°C (221°F)

Materials of Construction

Cell Body: PES (polyethersulfone)

 

Electrodes: see ‘Electrode Material’

 

 

Cable Options

Leadwire: PVC insulated 18 gage cable

 

20 and 50 feet lengths available

 

Quick Disconnect Option

 

Universal Head (Aluminum)

 

 

Approvals

Manufactured to comply with ASME boiler and

 

pressure vessel code Section III, Div.1, UG-101

 

CRN #0F11607.5C

 

 

Insertion

3/4” NPT male, Schedule 40 and 80 pipes

 

 

Insertion Depth

3-1/2” (89 mm) for 1, 10 and 0.01 cell constants

 

from solution end of 3/4" MNPT

 

2-1/2” (64mm) for 0.1 cell constant

 

 

2.2 Specifications for 055919 Flow Chamber

Parameter

Description

 

 

Maximum Flow

2 gpm @ 40psig and atmospheric discharge

 

 

Maximum Pressure

200 psig @ 25°C

 

 

Maximum Temperature

140°C (284°F) at atmospheric pressure

 

 

Dimensions

See Figure 3-1

 

 

Materials of Construction

Polyethersulfone (PES)

 

 

4

Installation and Maintenance Manual

10/07

Page 10
Image 10
Honeywell manual Specifications for 04973 Series, Specifications for 055919 Flow Chamber