Electrical Connections

4. Electrical Connections

4.1 Overview

The terminal board connections for the various Honeywell measuring instruments are given in the appropriate Figures in this section.

To avoid the possibility of AC pickup in the cell leads, separate them from all AC line-voltage wiring or run them in a separate grounded conduit.

ATTENTION

Do not use shielded cable except where shown in the following figures.

4.2 Instrument Wiring for 4973 Cells with Integral Cable

4.2.1 Model 4973 Series with Integral Cable to UDA2182 Analyzer

NOTES

1.FORPUREWATERSAMPLESINNON-CONDUCTIVE(PLASTIC,GLASS,ETC.) PIPING,GROUND THEBLACKCELLELECTRODELEADNEARTHECELL.ALTERNATIVELY,CONNECTTOTHEUDA GROUNDSCREW ASSHOWNDOTTED.DONOTGROUND10,25,OR50CONSTANTCELLS.

2.FORCELLLEADSBLACKANDWHITE,USE16TO22AWGCABLE,SHIELDEDTWISTEDPAIR, WITH30pFMAX.CAPACITANCEBETWEENCONDUCTORS.CONNECTSHIELDTOTERMINAL “10”.FORMODEL4973INSTALLATIONSWITHHONEYWELL31079198SSFLOW CHAMBER,USE 22AWGMINIMUMCOAXIALCABLETYPERG59U(BELDEN9259OREQUIV.).CONNECTCOAX SHIELDTOTERMINALS“A”AND“10”.CONNECTCOAXCONDUCTORTOTERMINALS“C”AND“7”.

3.FORCOMPENSATORLEADSBANDD,USE16TO22AWG,TWOCONDUCTORCABLE.

4.CELLTOANALYZERCABLESARECONSIDEREDLOW LEVEL.RUNSEPARATEFROMHIGH LEVELWIRING.

DRAWING 50012874

Figure 4-1 Installation Diagram, 4973 Cells, with junction box head connected to UDA2182

Analyzer

10

Installation and Maintenance Manual

10/07

Page 16
Image 16
Honeywell 4973 Series manual Electrical Connections, Instrument Wiring for 4973 Cells with Integral Cable