Installation

3.3 In-Line Type Mounting

In addition to the General Requirements outlined above, note the following with regard to insertion-type mounting:

Make certain the liquid head is above the cell location during measurement. A vertical insertion (from above) or a horizontal insertion can be used.

Allow at least one-half inch clearance beyond the end of the cell and 1/8 to 3/16 inch radius clearance to permit circulation of the solution.

It is usually best to have the solution flow up into the end of the cell since it is less likely to result in clogging by solids settling in the cell channels.

To be sure that a representative sample is being measured at all times, the solution must continuously move through the cell channels. In a rapidly moving solution, the assembly may be mounted so that the existing circulation forces the solution through the channels. When measurements are made in quiescent solutions, artificial means must be provided to force the solution through the cell. In some cases, this may be accomplished by moving the cell up and down.

Installation

Tighten the cell into a 3/4” NPT threaded opening (do not exceed a tightening torque greater than that indicated in Section 3.5) using a Teflon thread compound (preferably Teflon tape).

3.4 Flow-Type Mounting

In addition to the General Requirements outlined in Section 3.1, note the following with regard to flow- type mounting:

When mounting the cell in a pipe tee or flow chamber such as shown in Figure 1-3, have the solution enter the tee from below and exit to the side or from side and exit top. Be sure the electrodes are always as far as possible below the horizontal pipe run so that they are always covered to insure flooding of the cell under all conditions; otherwise, the conductivity reading may indicate a value that is lower than expected.

In general, the cell should be mounted so that the sample will flow through the channel toward the mounting end of the cell, exiting through the other channel hole or through the outer electrode holes. See Figure 1-1.

Locate the cell on the pressure side, not the vacuum side, of pumps. See Figure 1-2.

Avoid a horizontal cell mounting having the flow channel, see Figure 1-1, opposite to the flow exit of the pipe line, especially for the 1 and 10 constant cells. If necessary, refer to Section 5.5.

The 3/4” tee arrangement, Figure 1-3, assures that the cell is immersed well into the flow stream to obtain a representative sample. The tee is not supplied.

Installation

1.Tighten the cell into a 3/4” pipe tee (do not exceed a tightening torque greater than that indicated in Section 3.5).

2.If the flow-cell housing is used, assemble the cell and housing and install it in the process flow line or in a bypass line as indicated in Figure 1-2.

3.To avoid cracking the 055919 flow chamber, use Teflon tape on cell threads and tighten cell only enough to prevent leakage.

6

Installation and Maintenance Manual

10/07

Page 12
Image 12
Honeywell 4973 Series manual In-Line Type Mounting, Flow-Type Mounting