TB 9-6625-2049-35

 

 

Table 3. Amplitude

 

 

 

 

 

Test instrument settings

 

Multimeter indications

 

ENTRY AMPTD

ENTRY FREQ

 

 

(V ac)

 

3.536

V RMS

1 kHz

 

3.495

 

3.577

 

3.536

V RMS

100 kHz

 

3.495

 

3.577

 

1.061

V RMS

100 kHz

 

1.048

 

1.073

 

1.061

V RMS

1 kHz

 

1.048

 

1.073

 

1.061

V RMS

100 Hz

 

1.048

 

1.073

 

.3536

V RMS1

100 Hz

 

0.3416

 

0.3660

 

.3536

V RMS

1 kHz

 

0.3416

 

0.3660

 

.3536

V RMS

100 kHz

 

0.3416

 

0.3660

 

1Press ENTRY DC OFFSET key and enter 1 mV using DATA keys.

(5)Press keys and enter values using DATA keys as listed in (a) through (g) below:

(a)ENTRY DC OFFSET.

(b)0 mV.

(c)ENTRY FREQ.

(d)1 kHz.

(e)ENTRY AMPTD.

(f).707 V RMS.

(g)AMPTD CAL.

(6)Press MODIFY keys for a multimeter indication as close as possible to 0.707 V

ac.

(7)Enter 100 kHz using DATA keys. Multimeter will indicate between 0.683 and 0.732 V ac. Record multimeter indication.

(8)Disconnect multimeter and 50Ω feedthrough termination from TI.

NOTE

If necessary, ZERO and CALIBRATE measuring receiver and sensor module.

(9)Connect measuring receiver sensor module to TI SIGNAL.

(10)Set measuring receiver to measure volts at .1 MHz.

(11)Press ENTRY AMPTD key.

(12)Press MODIFY keys for a measuring receiver indication as close as possible to indication recorded in (7) above.

(13)Press ENTRY FREQ key and enter 500 kHz using DATA keys.

(14)Set measuring receiver to measure volts at .5 MHz. Measuring receiver will indicate between 0.683 and 0.732 V.

(15)Repeat technique of (13) and (14) above using TI and measuring receiver settings listed in table 4. Measuring receiver will indicate between 0.683 and 0.732 V.

15