HP 3325B, 3325A manual Adjustments, SIG Input

Models: 3325B 3325A

1 23
Download 23 pages 46.39 Kb
Page 12
Image 12

TB 9-6625-2049-35

(15) Set frequency counter controls to measure frequency. If frequency counter does not indicate between 59.99970 and 60.00030 MHz, perform b below.

b. Adjustments

(1) Disconnect TI rear panel 10 MHz OVEN OUTPUT from EXT REF IN option 001 only.

(2)Connect frequency counter CHANNEL A to TI rear panel AUX 21-60 MHz.

(3)Press keys and enter values using DATA keys as listed in (a) through (c) below:

(a)FUNCTION sine wave.

(b)ENTRY FREQ.

(c)60 MHz.

(4)Adjust REF R30 (fig. 3) until frequency counter indicates 60.000000 MHz (R).

Figure 3. Test instrument - bottom view.

(5)Disconnect frequency counter from TI rear panel AUX 21-60 MHz.

(6)Connect frequency counter CHANNEL A to TI SIGNAL.

(7)Enter 20 MHz using DATA keys. Frequency counter will indicate between 19.99990 and 20.00010 MHz.

(8)Disconnect frequency counter from TI.

(9)Perform (10) through (14) below for option 001 only.

(10)Connect time/frequency workstation OUTPUT 1 MHz to frequency difference meter REF INPUT.

(11)Connect TI rear panel 10 MHz OVEN OUTPUT to frequency difference meter

SIG INPUT.