SECTION 4—SPECIFICATIONS

SECTION 4—SPECIFICATIONS

Electrical Requirements:

120 VAC ± 10% (132 VAC/108 VAC), 60

 

Hz

 

 

Rated Current Input:

Platinum 5 and XL: 4.3 A

 

 

 

Platinum 10: 4.7 A

 

 

Sound Level:

Platinum 5 and XL: 50 dB Average

 

 

 

Platinum 10: 58 dB Average

 

 

Altitude:

Platinum 5 and XL - Up to 6,000 ft. (1828

 

meters) above sea level without degradation

 

of concentration levels. From 6,000 ft. (1828

 

meters) to 13,129 ft. (4000 meters) below

 

90% efficiency

 

 

 

Platinum 10 - Up to 4,000 ft. (1230 meters)

 

above sea level without degradation of con-

 

centration levels. From 4,000 ft. (1230

 

meters) to 13,129 ft. (4000 meters) below

 

90% efficiency

 

 

*Oxygen Output

All 5LXO2/5LX models 95.6% to 87% at 0.5

Concentration Levels:

to 5 L/min. and all 10LXO2/10LX models

 

95.6% to 87% at 2 L to 10 L/min.

 

 

Maximum Outlet Pressure:

Platinum 5 and XL: 5 psi ± 0.5 psi. (34.5 kPa

 

± 3.45 kPa.)

 

 

 

Platinum 10: 9 psi ± 0.5 psi. (62.0 kPa ± 3.45

 

kPa.)

 

 

*NOTE: Stated concentration levels achieved after initial warm-up period (approximately 30 minutes).

Part No 1118353

13

PlatinumSeries

Page 13
Image 13
Invacare IRC5LXO2, IRC10LXO2 manual Specifications