SECTION 7—TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

SOLUTION:

 

 

 

Alarm:

LOW FLOW ALARM

1a. Inspect for kinks or

RAPID

1a. Kinked or blocked

blockages. Correct, clean or

 

tubing, cannula or humid-

replace item. Once

Beep....Beep...

ifier.

corrected, turn power OFF

Beep...Beep

 

for 60 seconds and then turn

 

1b. Flowmeter set at 0.5

power back ON.

 

L/min. for Platinum 5 and

 

 

1 L/min. for Platinum 10

1b. Reset flowmeter to pre-

 

or less.

scribed flowrate. If prescribed

 

 

flowrate is below 0.5 L/min.,

 

 

use the Pediatric Flowmeter

 

 

IRCPF16 (Only with Platinum

 

 

5 and XL).

 

 

 