SECTION 7—TROUBLESHOOTING GUIDE

SECTION 7—TROUBLESHOOTING GUIDE

SYMPTOM:

PROBABLE CAUSE:

 

SOLUTION:

 

 

 

 

Alarm:

Main Power Loss:

1.

Insert plug into outlet.

 

 

 

 

 

SHORT BEEPS,

1.

Power cord NOT

2.

Inspect house circuit

LONG PAUSE

plugged in.

breakers or fuses. If problem

 

 

 

recurs, use a different outlet.

 

 

 

 

 

Concentrator

2.

No power at outlet.

3.

Call Service Provider.

NOT operating,

 

 

 

 

power switch

 

 

 

 

ON.

 

 

 

 

 

 

 

 

 

Beep......Beep

3.

Tripped circuit

 

 

 

breaker.

 

 

 

 

 

Alarm:

System Failure:

1a. Remove and clean cabinet

CONTINUOUS

1.

Unit overheating due

filters.

 

to blocked air intake.

 

 

Concentrator

 

 

1b. Move oxygen concentra-

NOT operating,

2.

Insufficient power at

tor at least three inches away

Power Switch

outlet.

from walls, draperies or furni-

ON.

 

 

ture.

 

3.

Internal repairs

 

 

Beeeeeeeep....

required

2.

DO NOT use extension

 

 

 

cords. Move to another elec-

 

 

 

trical outlet or circuit.

 

 

 

3.

Call Service Provider.

 

 

 

 

 

Part No 1118353

35

PlatinumSeries

Page 35
Image 35
Invacare IRC10LXO2, IRC5LXO2 manual Troubleshooting Guide, Solution