SECTION 5—OPERATING INSTRUCTIONS

Flow Knob

Ball

FIGURE 5.6 Flowrate

SensO2 Oxygen Purity Indicator (If Your Unit Features the O2 Sensor Feature)

This feature monitors the purity of oxygen generated by the oxygen concentrator. If purity falls below factory preset standards, indicator lights on the control panel will illuminate.

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm-up time (approximately 30 min.) while waiting for the O2 purity to reach maximum.

When the unit is turned on, the GREEN light will come on (SYSTEM OK/O2 greater than 85%). After five (5) minutes, the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values.

Explanation of Oxygen Purity Indicator Lights

NOTE: For this procedure, refer to FIGURE 5.7 on page 30.

GREEN light (O2) - Normal Operation.