SECTION 7—TROUBLESHOOTING GUIDE
SYMPTOM: | PROBABLE CAUSE: | SOLUTION: | |
|
|
|
|
YELLOW or RED | 1. | Low oxygen purity.* | 1. Clean or Replace filters. |
LIGHT Illumi- |
|
|
|
nated. | 2. | Kinked or blocked tub- | 2. Inspect for kinks or block- |
| ing, cannula or humidi- | ages. Correct, clean or | |
| fier.* | replace item. Once cor- | |
|
|
| rected, turn power OFF for |
Alarm: | 3. | Flowmeter set at 0.5 | 60 seconds and then turn |
CONTINUOUS | L/min. for Platinum 5 and | power back ON. | |
On red light only. | 1 L/min. for Platinum 10 |
| |
| or less.* | 3. Check flowmeter is set to | |
|
|
| 1.0 L/min. or more. (Refer to |
| 4. | Unit overheating due | Recommended Guidelines for |
| to blocked air intake. | Optimum Performance on | |
|
|
| page 24. |
| 5. | Insufficient power at |
|
* Only applies to | outlet. | 4.a. Remove and clean cabinet | |
5LXO2 and |
|
| filters. |
10LXO2 SensO2 | 6. | Internal repairs | 4b. Move concentrator at |
models. | required. | ||
|
| least three inches from walls, | |
|
|
| |
|
|
| draperies, and furniture. |
|
|
| 5. DO NOT use extension |
|
|
| cords. Move to another elec- |
|
|
| trical outlet or circuit. |
|
|
| 6. Call Service Provider. |
|
|
|
|
GREEN LIGHT | 1. | Internal repairs | 1. Call Service Provider. |
with YELLOW | required. |
| |
Light Flashing. |
|
|
|
|
|
|
|
Platinum™Series | 36 | Part No 1118353 |