PlatinumSeries

34

Part No 1118353

FIGURE 6.2 Preventive Maintenance Record

PREVENTIVE MAINTENANCE RECORD

INVACARE PLATINUM OXYGEN CONCENTRATOR

ON EACH INSPECTION

Record Date of Service

Record Elapsed Hours On Hour Meter

Clean Cabinet Filter(s)

Check Prescribed L/min. Flow Rate

EVERY 180 DAYS FOR IRC5LX (EVERY 90 DAYS FOR IRC10LX)

Check Oxygen Concentration

(UNITS WITHOUT SensO2)

ANNUALLY, DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS

Clean/Replace Cabinet Filter(s)

Check Outlet HEPA Filter*

Check Compressor Inlet Filter*

Check Oxygen Concentration

(SensO2 MODELS)

Check Power Loss Alarm

AS REQUIRED

Clean Heat Exchanger

Replace Exhaust Muffler

Rebuild Top End of Compressor

*Refer to Preventive Maintenance section of Service Manual.

MODEL NO. I RC

SERIAL NO.

WARNING

DO NOT overfill humidifier bottle.

SECTION 6—MAINTENANCE

Page 34
Image 34
Invacare IRC5LXO2, IRC10LXO2 manual Preventive Maintenance Record