SECTION
5LXO2 INDICATORS:
Label
Symbol
O2 Purity
Indicator Lights
O2
SYSTEM FAILURE
Below 73%
73% to 85%
SYSTEM OKAY
Over 85%
RED - Continuous Audible
Alarm
Compressor Shutdown
YELLOW
GREEN
5LX INDICATORS:
| Label | Status | Indicator Lights |
| |||
| Symbol |
|
| ||||
|
|
|
|
|
|
| |
|
|
|
|
|
|
| |
|
|
| SYSTEM FAILURE |
| RED |
| |
| I/O |
| SYSTEM OKAY |
| GREEN |
| |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
Flow Range: |
| Platinum 5 and XL - 0.5 to 5 L/min. For | |||||
|
|
|
| flowrates less than 1 L/min., we recommend | |||
|
|
|
| the use of the Invacare Pediatric Flowmeter | |||
|
|
|
| IRCPF16 Accessory. | |||
|
|
|
|
|
|
| |
|
|
|
| Platinum 10 - 2 L/min. to 10 L/min | |||
|
|
|
|
|
|
| |
Low Flow Alarm: |
| Platinum 5 - 0 L/min. to 0.5 L/min., | |||||
|
|
|
| Platinum 10 - 0 L/min. to 1 L/min. | |||
|
|
|
| Rapid Audible Alarm Beeping (No | |||
|
|
|
| Accessories Connected) | |||
|
|
|
|
|
|
| |
Average Power |
| Platinum 5 and XL: 400W | |||||
Consumption: |
| Platinum 10: 585W | |||||
|
|
|
|
|
| ||
Pressure Relief Mechanism | 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa) | ||||||
Operational at: |
|
|
|
|
| ||
|
|
|
|
|
|
| |
Change in maximum recom- | 0.7 L/min. |
|
| ||||
mended flow when back pres- |
|
|
|
| |||
sure of 7kPa is applied: |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
Platinum™Series | 14 | Part No 1118353 |