SECTION 3—OPERATION

Stand Up Patient Lift 12 Part No. 1078984
SECTION 3— OPERATION
WARNING

DO NOT attempt to transfer a patient without approval of the patient’s physician, nurse, or medical

assistant. Thoroughly read the instructions in this owner’s manual, observe a trained team of experts

performing the lifting procedures and then perform the entire lift procedure several times with

proper supervision and a capable individual acting as a patient.

The legs of the stand up lift MUST be in the maximum open position for optimum stability and

safety. If the patient is in a sling and it becomes necessary to move through a narrow passage, close

the legs of the stand up lift only as long as it takes to move through the passage. When the stand up

lift is through the passage, return the legs to the maximum open position.

NOTE:Theuseofthepatientliftbyoneassistantshouldbebasedon
theevaluationofthehealthcareprofessionalforeachindividualcase.

Locking/Unlocking the Rear Casters

•RefertoDetail“A”ofFigure3.1.

Raising/Lowering the Patient Lift

NOTE:Forthisprocedure,refertoDetail“B”.
•ToRaisethePatientLift:
Presstheup() buttononthehandcontrol.
NOTE:Ifthepatientliftisraisedtothehighestlevel,itmaybe
necessarytopulldowngentlyontheliftarmsbeforethemastwill
begintolower.Pulldowngentlyonbotharmsatthesametimeto
avoidmakingtheliftunstable.
•ToLowerthePatientLift:
Pressthedownbutton() onthehandcontrol.

Closing/Opening the Legs

•ToClosetheLegs:
PulltheshifterhandleOUTandawayfromthestanduplift
andthentoyourLEFTuntilitLOCKSinthenotchofthe
bracket.
NOTE:Leftisdeterminedbystandingbehindthestanduplift
facingtowardsthefrontcasters.
•ToOpentheLegs:
PulltheshifterhandleOUTandawayfromthestanduplift
andthentoyourRIGHTuntilitLOCKSinthenotchofthe
bracket.
NOTE:Rightisdeterminedbystandingbehindthestanduplift
facingtowardsthefrontcasters.

Using the Emergency Stop

NOTE:Forthisprocedure,refertoDetail“C”.
•PushtheREDemergencybuttonintostoptheliftarms
fromraisingorlowering.
•Toreset,rotatetheemergencybuttonclockwise.
FIGURE 3.1 Operating the Patient Lift
DETAIL “B” - HAND
CONTROL DETAILS
DETAIL “C” - EMERGENCY STOP
Down Button
Up Button
Hand Control
Emergency Button
Push in to stop,
turn clockwise
to reset.
Step Here
to Lock
Step Here
to Unlock
Locking
Lever
LOCKED
UNLOCKED
DETAIL “A”