IRIS P522 manual Adjusting Set Points, PLICATION TABLES are, Gain potentiometers POTS full

Models: P522

1 40
Download 40 pages 16.44 Kb
Page 17
Image 17
PLICATION TABLES (Figure 11) are

IRIS MODEL P522 APPLICATION MANUAL

The above settings will be satisfactory for most applications, but are used only as a starting point. So many variables can affect the readings that it is impossible to catalogue all the settings, and each viewing head has its own characteristics for the different types of fuels and burn- ers. We recommend that you become knowledgeable about the characteristics of the different viewing head models available before applying them to a par- ticular application. Detailed descriptions of the viewing heads are provided in this manual, and the VIEWING HEAD AP-

There must be a definitive ratio between the read- ings for burner ON and burner OFF. This ratio can be defined by using the lowest reading for burner ON from both of the load conditions, and the high- est reading for burner OFF from both of the load conditions. If the readings are 2:1 or more (BNR ON to BNR OFF) then there should be no problem discriminating between burners.

A good balance for the set points would be:

ON SET POINT = .75A + .25B

OFF SET POINT = .25A + .75B

PLICATION TABLES (Figure 11) are

where A is the lowest burner ON reading, and B is

a useful summary of the viewing heads

the highest burner OFF reading.

and their response to different fuels.

Actually, a smaller ratio will work, as long as the

B. The initial settings for the flicker type

lowest BNR ON reading and the highest BNR OFF

viewing heads (S509, S511, S512) should

reading never deteriorates to the point the former is

be as follows:

equal to or less than the latter. In other words, A

1. Set filter switch to LL position.

must always be greater than B. Using a safety factor

of 2:1 will allow for flame signal changes that prob-

 

2. Gain potentiometers (POTS) full

ably will occur over a period of time.

clockwise (these are 25-turn potenti-

A. FOR FLICKER TYPE VIEWING

ometers – a slight clicking sound will

HEADS: If unable to achieve discrimi-

be heard when the POT is maximum

nation from background signals (for mod-

clockwise). There are two POTS on

els S509, S511 and S512 viewing heads),

the S509: one for the Si channel, and

remove the viewing head housing from

one for the PbS channel. If the fuel is

the flange and place the filter switch to

natural gas, turn the Si channel off

the next highest position L. Reinstall the

(POT full counter clockwise). Refer

viewing head and take another set of read-

to section S509 VIEW HEAD.

ings. Note: the filter switch should be

 

C. There are no initial settings for the S506

moved only one position at a time for

UV viewing head. It is important that

each set of readings. Refer to section

this viewing head be properly aimed to

MULTIBURNER REQUIREMENTS.

pick up the maximum signal, as described

B. FOR THE S509 VIEWING HEAD: If

in the previous section.

you are still unable to achieve satisfac-

 

ADJUSTING SET POINTS

tory flame OFF operation, remove the

Before adjusting the set points, take readings with

viewing head and turn the Si channel off

by turning the Si POT at least 30 turns

the P522 on the two worst-case conditions described

counter clockwise. The signal level will

(see section SETUP AND ADJUSTMENT PRO-

now be lower because only the PbS de-

CEDURES). Write down the readings for the burner

tector is being used, so the set points must

ON and OFF (with other burners on) under low load

be re-evaluated. If you are using an S511

conditions, and then the readings for the burner ON

viewing head and not achieving success-

and OFF under high load conditions.

 

 

 

 

Page 17

Page 17
Image 17
IRIS P522 manual Adjusting Set Points, PLICATION TABLES are, Gain potentiometers POTS full, A. For Flicker Type Viewing