secondary readings are not achieved,

 

 

replace transformer.

 

Temperature display

If proper voltage is present at terminals

 

 

1,2, and 3 on temperature display, but

 

 

but the display is not lighted, recheck

 

 

all connections for tightness. If

 

 

temperature display is still not operating,

 

 

replace the temperature display.

Temperature display

Thermocouple

Connect the thermocouple of a

Inaccurate, erratic

 

pyrometer to the thermocouple on

Or inoperative

 

the temperature control board. (be sure

 

 

to maintain proper polarity of

 

 

thermocouple leads—see adjustment

 

 

section for proper procedure.) The

 

 

pyrometer will indicate oven temperature

 

 

display. Measure D.C. millivolts of

 

 

thermocouple at the temperature display.

 

 

See thermocouple chart in section D for

 

 

proper readings. Replace thermocouple

 

 

if proper readings are not achieved. If

 

 

readings are correct on the thermocouple,

 

 

replace temperature display.