ERROR INDICATORS

Indication:

Cuff pressure reaches 330mmHg

Cause:

Twisted bladder inside off the cuff

Correction:

Make sure that the bladder lays flat within the cuff.

 

Carefully read and repeat the above mentioned steps.

Indication:

Displays "Err", cuff pressure releases and "0"

 

pressure displays.

Cause:

Excessive arm movement or talking during

 

measurement.

Correction:

Press Start/Stop button to measure again while

 

keeping still.

 

Batteries are used. Please change the two batteries

 

with new ones.

CARE AND MAINTENANCE

-Do not clean the monitor with volatile liquids. The monitor should be cleaned with a soft, dry cloth.

-Take out the batteries if the unit is not being used for a long period of time.

Model

MS-900VX

Type

Oscillometric

Display

LCD: size: 26 x 26 mm

Measuring range

0 – 300 mmHg / 0-40 kPa

 

(blood pressure)

 

30 – 190 pulse beats pr. minute

Accuracy

± 3mmHg / ± 0.4 kPa (blood pressure)

 

± 5% (pulse)

Pressure

Fuzzy logic

Deflation system

Electro-mechanical release valve

Cuff

Velcro with mechanical closing

Power source

2 pcs. AAA batteries – Alkaline

 

(Not incl.)

ENVIRONMENTAL TIP

An electronic appliance should, when it is no longer capable of functioning be disposed with least possible environmental damage. The appliance should be disposed according to the local regulations in your municipality, in most cases you can discharge the appliance at your local recycling center

THE WARRENTY DO NOT COVER

-If the above points have not been observed.

-If the appliance has not been properly maintained, if force has been used against it or if it has been damaged in any other way.

-If the appliance has been repaired or modified or changed in any way or by any person not properly authorized.

45

Page 45
Image 45
Melissa 630-010 manual Error Indicators, Care and Maintenance, Environmental TIP, Warrenty do not Cover