Basic Maintenance

Basic Maintenance

Opening the Meter CaseWarning

To avoid shock, remove the test leads and any input signals before opening the case or replacing the battery or fuses.

Opening the Meter Case

Caution

To avoid unintentional shock circuit, always place the uncovered Meter assembly on a protective surface. When the case of the Meter is open, circuit connections are exposed.

To open the Meter case, refer to Figure 1 and do the following:

1.Disconnect test leads from any live source, turn rotary switch to OFF, and remove the test leads from the front terminals.

2.Remove the battery door by using a flat-blade screwdriver to turn the battery door screws turn counter-clockwise.

3.The bottom case is secured to the top case by four screws. Using a phillips-head screwdriver to remove the four screws.

Replacing the Battery

The meter is powered by 2 AA 1.5V batteries.

To replace the battery, refer to Figure

10