Testing the Voltage Function

To verify accuracy in the AC and DC voltage ranges, do the following:

1.Turn the rotary switch to “VTesting the Voltage Function” position.

2.Connect the Calibrator to the VΩ and COM inputs on the Meter.

3.Set the Calibrator for the voltage and frequency from step 1 to 12 in Table 1.

4.Compare the reading on the Meter display with the display reading shown in Table 1.

5.If the display reading falls outside of the range shown in Table 1, the Meter does not meet specification.

Table 1 AC Voltage Test:

Step

Input

Frequency

Reading

1

500mV

50Hz

494.7 to 505.3

 

 

 

 

2

500mV

500Hz

494.7 to 505.3

 

 

 

 

3

500mV

1kHz

489.7 to 510.3

 

 

 

 

4

5.000V

50Hz

4.947 to 5.053

 

 

 

 

5

5.000V

500Hz

4.947 to 5.053

 

 

 

 

6

5.000V

1kHz

4.897 to 5.103

 

 

 

 

7

50.00V

50Hz

49.47 to 50.53

 

 

 

 

8

50.00V

500Hz

49.47 to 50.53

 

 

 

 

9

50.00V

1kHz

48.97 to 51.03

 

 

 

 

10

500.0V

50Hz

494.7 to 505.3

 

 

 

 

11

500.0V

500Hz

494.7 to 505.3

 

 

 

 

12

500.0V

1kHz

489.7 to 510.3

 

 

 

 

6.Turn the rotary switch to “VManual background” position.

7.Set the calibration for the voltage from step 1 to 4 in Table 2.

8.Compare the reading on the Meter display with the display reading shown in Table 2.

9.If the display reading falls outside of the range shown in Table 2, the meter does not meet specification.

Table 2 DC Voltage Test:

Step

Input

Reading

1

500mV

497.3 to 502.7

 

 

 

2

5.000V

4.973 to 5.027

 

 

 

3

50.00V

49.73 to 50.27

 

 

 

4

500.0V

497.3 to 502.7

 

 

 

13

Page 15
Image 15
Milwaukee 2216-20 manual Testing the Voltage Function