(E)DCA Calibration (VR7)

1.Rotate the rotary switch to DCA Function.

2.Connect the output of a DCA calibrator to V and COM input terminals of the meter.

3.Output DCA 5.000A from the calibrator.

4.Adjust VR7 until the display shows 5.000A ± 0.001A.

5.Turn off the output of the calibrator and disconnect the meter from the calibrator.

(F)Resistance (6kΩ Range) Calibration (VR6)

1.Rotate the rotary switch to the resistance function.

2.Connect the output of a resistance calibrator to V and COM input terminals of the meter.

3.Output resistance 5.000kΩ from the calibrator.

4.Adjust the VR6 until the display shows 5.000kΩ ± 1Ω.

5.Turn off the output of the calibrator and disconnect the meter from the calibrator.

(G)Capacitance Calibration (VR1)

1.Rotate the rotary switch to Capacitance Function.

2.Connect the output of a capacitance calibrator to V and COM input terminals of the meter.

3.Output capacitance 100.0μF from the calibrator.

4.Adjust VR1 until the display shows 100.0μF ± 0.1μF.

5.Turn off the output of the calibrator and disconnect the meter from the calibrator.

(H)Assemble the bottom case

1.Assemble the bottom case.

2.Tighten the screw.