AppendixA Specifications
NI5911 User Manual A-6 ni.com
TriggeringSystems
Modes.....................................................Above threshold, below
threshold, between thresholds,
outside thresholds
Source.....................................................CH0, RTSI<0..6>, PFI 1,2
Slope.......................................................Rising/falling
Hysteresis................................................Full-scalevoltage/n,wherenis
between 1 and 170; full-scale
voltage on TRIG is fixed to ±5 V
(without external attenuation)
Coupling.................................................AC/DC on CH0, TRIG
Pretriggerdepth ......................................1 to 16 million samples
Posttriggerdepth .....................................1 to 16 million samples
Holdoffby time ......................................40 ns 171.85 s in increments
of 40 ns
Sensitivity...............................................170 steps in full-scale voltage
range
TRIG input range....................................±5 V (without external
attenuation)
TRIG input impedance ...........................1 M± 1% in parallel
with 30 pF ± 15 pF
TRIG input protection ............................±42V [(DC + peak AC) < 10 kHz,
without external attenuation]