AppendixA Specifications
NI5911 User Manual A-8 ni.com
Calibration
Internal....................................................Internal calibration is done upon
softwarecommand. The
calibration involves gain, offset
and linearitycorrection for all
input ranges and input modes.
Interval.............................................1week, or any time temperature
changesbeyond ±5 °C. Hardware
detects temperature variations
beyond calibration limits, which
can also be queried by software.
External...................................................Internalreference requires
recalibration
Interval.............................................1year
Warm-uptime .........................................1 minute