Chapter 2 Configuration and Installation
© National Instruments Corporation 2-3 NI 7350 User Manual
Remove power from signal lines before connecting them to or
disconnecting them from the device.
Operate the device at or below the measurement category1 marked on the
hardware label. Measurement circuits are subjected to working voltages2
and transient stresses (overvoltage) from the circuit towhich they are
connected during measurement or test. Measurement categories establish
standard impulse withstand voltage levels that commonly occur in
electrical distribution systems. The following is a description of
measurement categories:
• Measurement Category I is for measurements performed on circuits
not directly connected to the electrical distribution system referred to
as MAINS3 voltage. This category is for measurements of voltages
from specially protected secondary circuits. Such voltage
measurements include signal levels, special equipment, limited-energy
parts of equipment, circuits powered by regulated low-voltage sources,
and electronics.
• Measurement Category II is for measurements performed on circuits
directly connected to the electrical distribution system. This category
refers to local-level electrical distribution, such as that provided
bya standard wall outlet (for example, 115 AC voltage for U.S. or
230 ACvoltage for Europe). Examples of Installation Category II are
measurements performed on household appliances, portable tools,
and similar devices/modules.
• Measurement Category III is for measurements performed in the
building installation at the distribution level. This category refers to
measurements on hard-wired equipment such as equipment in fixed
installations, distribution boards, and circuit breakers. Other examples
are wiring, including cables, bus bars, junction boxes, switches, socket
outlets in the fixed installation, and stationary motors with permanent
connections to fixed installations.
• Measurement Category IV is for measurements performed at the
primary electrical supply installation (<1,000 V). Examples include
electricity meters and measurements on primary overcurrent
protection devices and on ripple control units.
1 Measurement categories, also referred to as installation categories, are defined in electrical safety standard IEC 61010-1.
2 Working voltage is the highest rms value of an AC or DC voltage that can occur across any particular insulation.
3 MAINS is defined as a hazardous live electrical supply system that powers equipment. Suitably rated measuring circuits may
be connected to the MAINS for measuring purposes.