Appendix A Specifications
NI VXIpc-882 User Manual A-2 ni.com
Operating Environment
Ambient temperature range....................5 to 50 °C
(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2.)
Relative humidity range..........................10 to 90%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Altitude...................................................2,000 m (at 25 °C ambient
temperature)
Pollution Degree.....................................2
For indoor or enclosed area use only.
Module cooling requirements.................For 18 °C temperature rise,
2.6 liters/sec airflow,
0.03 mm H2O pressure across
module, per VXI-8, Cooling
Characterization Methodology
Specification
Caution Clean the NI VXIpc-882 with a soft nonmetallic brush. Make sure that the device
is completely dry and free from contaminants before returning it to service.
Storage Environment
Ambient temperature range....................–20 to 70 °C
(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2.)
Relative humidity range..........................5 to 95%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)