SCXI-1120 Cabling Appendix E
SCXI-1120 User Manual E-4 © National Instruments Corporation
Lab boards full access to the digital control lines and analog signals (except DTEMP), but the Lab
boards cannot scan channels in Multiplexed mode. Leave jumper W1 in position A on the
SCXI-1341 and SCXI-1344. The SCXI-1120 does not use jumper W1. Table E-2 lists the
SCXI-1341 and SCXI-1344 pin translations.
Note: If you are using the Lab-PC+, configure the board for single-ended inputs.
Table E-2. SCXI-1341 and SCXI-1344 Pin Translations
Lab Board Pin Lab Board Signal SCXI-1120 Pin SCXI-1120 Signal
1 ACH0 3 MCH0+
2 ACH1 5 MCH1+
3 ACH2 7 MCH2+
4 ACH3 9 MCH3+
5 ACH4 11 MCH4+
6 ACH5 13 MCH5+
7 ACH6 15 MCH6+
8 ACH7 17 MCH7+
9 AIGND 1-2 AOGND
10 DAC0OUT 20 No Connect
11 AOGND 23 No Connect
12 DAC1OUT 21 No Connect
13, 50 DGND 24, 33 DIG GND
26 PB4 25 SERDATIN
27 PB5 27 DAQD*/A
28 PB6 29 SLOT0SEL*
29 PB7 37 SERCLK
31 PC1 26 SERDATOUT
32 PC2 28 No Connect
40 EXTCONV* 36 SCANCLK
43 OUTB1 46 No Connect
49 +5 V 34-35 No Connect
All other pins of the Lab board pinout are not sent to the SCXI-1120 rear signal connector.
SCXI-1341 and SCXI-1344 InstallationFollow these steps to install the SCXI-1341 or SCXI-1344:
1. Make sure that the computer and the SCXI chassis are turned off.
2. Install the SCXI module in the chassis.
3. Connect one end of the ribbon cable to the adapter board rear connector. This is the 50-pin
connector of the SCXI-1344 cable.
4. Plug the adapter board front connector to the module rear signal connector. Make sure a
corner of the adapter board enters the upper module guide of the chassis.
5. Screw the rear panel to the threaded strips in the rear of the chassis.