Configuration and Installation Chapter 2
SCXI-1120 User Manual 2-10 © National Instruments Corporation
Table 2-5. Gain Jumper Positions
Gain Setting Jumper Position
First-stage
1
10
50
100
D
C
B
A (Factory setting)
Second-stage
1
2
5
10
20
A
B
C
D (Factory setting)
E
Filter Jumpers
Two-stage filtering is also available on your SCXI-1120 module. The first stage is located in the
isolated section of the input channel, whereas the second stage is located in the nonisolated
section of your input channel. Two-stage filtering eliminates the noise generated by the isolation
amplifier, producing a higher signal-to-noise ratio. Furthermore, two filter bandwidths are
available, 10 kHz and 4 Hz.
Table 2-6. Filter Jumper Allocation
Input Channel First Filter Jumper Second Filter Jumper
Number 4 Hz
(Factory
Setting)
10 kHz 4 Hz
(Factory
Setting)
10 kHz
0 W17-A W17-B W25 W26
1 W18-A W18-B W27 W28
2 W19-A W19-B W29 W30
3 W20-A W20-B W31 W32
4 W21-A W21-B W33 W34
5 W22-A W22-B W35 W36
6 W23-A W23-B W37 W38
7 W24-A W24-B W39 W40
Your SCXI-1120 is shipped in the 4 Hz position. Remember to make sure that both stages are
set to the same bandwidth to ensure that the required bandwidth is achieved. Notice that one
jumper block is available for each filter stage.