Chapter 2 Configuration and Installation
© National Instruments Corporation 2-9 SCXI-1120 User Manual
Table 2-3. Jumper W41 Settings
Jumper Description Configuration
W41 Temperature sensor
accessed in MTS mode
(Factory setting)
•
•
•
3 2 1
W41 Temperature sensor
accessed in DTS mode,
data acquisition board
configured for NRSE or
RSE
•
•
•
3 2 1
Gain Jumpers
Each input channel has two gain stages. The first gain stage provides gains of 1, 10, 50, and 100,
and the second stage provides gains of 1, 2, 5, 10, and 20. Tables 2-4 and 2-5 show how to set
up the gain for each channel.
Table 2-4. Gain Jumper Allocation
Input Channel Number First Gain Jumper Second Gain Jumper
0W1W9
1 W2 W10
2 W3 W11
3 W4 W12
4 W5 W13
5 W6 W14
6 W7 W15
7 W8 W16
The board is shipped with the first-stage gain set to 100 (position A), and a second-stage gain set
to 10 (position D). To change the gain of your module, move the appropriate jumper on your
module to the position indicated in Tables 2-3 and 2-4. Refer to Figure 2-1, SCXI-1120 General
Parts Locator Diagram, and Figure 2-2, Detailed Parts Locator Diagram, for jumper locations
on your module.
To determine the overall gain of a given channel use the following formula:
Overall gain = First-stage gain x second-stage gain.